Quantitative analysis of hemodynamic changes induced by the discrepancy between the sizes of the flow diverter and parent artery

Title
Quantitative analysis of hemodynamic changes induced by the discrepancy between the sizes of the flow diverter and parent artery
Author
오제훈
Keywords
Flow diverter; Cerebral aneurysm; Metal coverage rate; Computational fuid dynamics; Hemodynamic changes; Size discrepancy
Issue Date
2024-05-09
Publisher
NATURE PORTFOLIO
Citation
SCIENTIFIC REPORTS, v. 14, article no. 10653, page. 1-8
Abstract
The efficacy of flow diverters is influenced by the strut configuration changes resulting from size discrepancies between the stent and the parent artery. This study aimed to quantitatively analyze the impact of size discrepancies between flow diverters and parent arteries on the flow diversion effects, using computational fluid dynamics. Four silicone models with varying parent artery sizes were developed. Real flow diverters were deployed in these models to assess stent configurations at the aneurysm neck. Virtual stents were generated based on these configurations for computational fluid dynamics analysis. The changes in the reduction rate of the hemodynamic parameters were quantified to evaluate the flow diversion effect. Implanting 4.0 mm flow diverters in aneurysm models with parent artery diameters of 3.0-4.5 mm, in 0.5 mm increments, revealed that a shift from oversized to undersized flow diverters led to an increase in the reduction rates of hemodynamic parameter, accompanied by enhanced metal coverage rate and pore density. However, the flow diversion effect observed transitioning from oversizing to matching was less pronounced when moving from matching to undersizing. This emphasizes the importance of proper sizing of flow diverters, considering the benefits of undersizing and not to exceed the threshold of advantages.
URI
https://www.nature.com/articles/s41598-024-61312-yhttps://repository.hanyang.ac.kr/handle/20.500.11754/190922
ISSN
2045-2322
DOI
https://doi.org/10.1038/s41598-024-61312-y
Appears in Collections:
COLLEGE OF ENGINEERING SCIENCES[E](공학대학) > MECHANICAL ENGINEERING(기계공학과) > Articles
Files in This Item:
There are no files associated with this item.
Export
RIS (EndNote)
XLS (Excel)
XML


qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.

BROWSE