An Assistive Image-Guided Surgical Robot System Using O-Arm Fluoroscopy for Pedicle Screw Insertion: Preliminary and Cadaveric Study

Title
An Assistive Image-Guided Surgical Robot System Using O-Arm Fluoroscopy for Pedicle Screw Insertion: Preliminary and Cadaveric Study
Author
이병주
Keywords
Image-guided surgery; Percutaneous pedicle screw insertion; Surgical navigation; Surgical planning; Surgical robot
Issue Date
2010-12
Publisher
Lippincott Williams & Wilkins Ltd.
Citation
Neurosurgery, v. 67, NO. 6, Page. 1757-1767
Abstract
BACKGROUND: The biplane fluoroscopy guided robot system (BFRS) was developed for surgical robotic systems, minimally invasive surgeries, and cooperative robotic systems, as well as enhanced surgical planning and navigation with preoperative and intraoperative image data. OBJECTIVE: To propose a novel surgical robot system for percutaneous pedicle screw insertion. METHODS: The BFRS consists of an O-shaped biplane fluoroscope (O-arm), a surgical planning and operating system, and an assistive robot. Each part of the BFRS has a role in conducting percutaneous pedicle screw placements. To evaluate BFRS accuracy, each part was analyzed, and to assess the safety and feasibility of percutaneous pedicle screw insertions with the BFRS, cadaveric studies involving 14 levels in the thoracic and lumbar spine regions were conducted on 2 cadavers. RESULTS: Errors in each part of the system and within the entire system were evaluated. The accuracy of generating coordinates using O-arm images was 0.30 +/- 0.15 mm. The robot demonstrated a duplication value of 4.97 mm RMS and an accuracy of 0.358 mm RMS. Total system error was 1.38 +/- 0.21 mm. The results of the cadaveric studies show that inserted pedicular screws were adequately located within the spine with no unexpected malpositioning of the screws. The axial angle difference between planned and postoperative data was 2.45 +/- 2.56 degrees, and the sagittal angle difference was 0.71 +/- 1.21 degrees. CONCLUSION: The BFRS might be helpful in improving the accuracy of percutaneous pedicular screw insertion procedures. In the future, we will attempt to improve the accuracy and reliability of the BFRS and to determine new clinical applications for the BFRS.
URI
https://www.proquest.com/docview/2307722180?accountid=11283https://repository.hanyang.ac.kr/handle/20.500.11754/181078
ISSN
0148-396X;1524-4040
DOI
10.1227/NEU.0b013e3181fa7e42
Appears in Collections:
COLLEGE OF ENGINEERING SCIENCES[E](공학대학) > ELECTRICAL ENGINEERING(전자공학부) > Articles
Files in This Item:
There are no files associated with this item.
Export
RIS (EndNote)
XLS (Excel)
XML


qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.

BROWSE