Compton-edge-based energy calibration of double-sided silicon strip detectors in Compton camera
- Title
- Compton-edge-based energy calibration of double-sided silicon strip detectors in Compton camera
- Author
- 김찬형
- Keywords
- Energy calibration; Compton edge; DSSD; Compton camera
- Issue Date
- 2011-05
- Publisher
- Elsevier Science B.V., Amsterdam.
- Citation
- Nuclear Instruments & Methods in Physics Research. Section A,Vol.633//SUP1, No.- [2011],S108-S110
- Abstract
- Abstract Accurate energy calibration of double-sided silicon strip detectors (DSSDs) is very important, but challenging for high-energy photons. In the present study, the calibration was improved by considering the Compton edge additionally to the existing low-energy calibration points. The result, indeed, was very encouraging. The energy-calibration errors were dramatically reduced, from, on average, 15.5% and 16.9% to 0.47% and 0.31% for the 356 ( 133 Ba) and 662keV ( 137 Cs) peaks, respectively. The imaging resolution of a double-scattering-type Compton camera using DSSDs as the scatterer detectors, for a 22 Na point-like source, also was improved, by ∼9%.
- URI
- http://www.sciencedirect.com/science/article/pii/S0168900210013288?via%3Dihubhttp://hdl.handle.net/20.500.11754/66616
- ISSN
- 0168-9002
- DOI
- 10.1016/j.nima.2010.06.138