Low-Tube-Voltage CT Urography Using Low-Concentration-Iodine Contrast Media and Iterative Reconstruction: A Multi-Institutional Randomized Controlled Trial. for Comparison with Conventional. CT Urography

Title
Low-Tube-Voltage CT Urography Using Low-Concentration-Iodine Contrast Media and Iterative Reconstruction: A Multi-Institutional Randomized Controlled Trial. for Comparison with Conventional. CT Urography
Author
김용수
Keywords
Computed tomography; Low dose; Urography; Contrast media; Double dose reduction
Issue Date
2018-12
Publisher
KOREAN RADIOLOGICAL SOC
Citation
KOREAN JOURNAL OF RADIOLOGY, v. 19, no. 6, page. 1119-1129
Abstract
Objective: To compare the image quality of low-tube-voltage and low-iodine-concentration-contrast-medium (LVLC) computed tomography urography (CTU) with iterative reconstruction (IR) with that of conventional CTU.Materials and Methods: This prospective, multi-institutional, randomized controlled trial was performed at 16 hospitals using CT scanners from various vendors. Patients were randomly assigned to the following groups: 1) the LVLC-CTU (80 kVp and 240 mgI/mL) with IR group and 2) the conventional CTU (120 kVp and 350 mgI/mL) with filtered-back projection group. The overall diagnostic acceptability, sharpness, and noise were assessed. Additionally, the mean attenuation, signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR), and figure of merit (FOM) in the urinary tract were evaluated.Results: The study included 299 patients (LVLC-CTU group: 150 patients; conventional CTU group: 149 patients). The LVLC-CTU group had a significantly lower effective radiation dose (5.73 +/- 4.04 vs. 8.43 +/- 4.38 mSv) compared to the conventional CTU group. LVLC-CTU showed at least standard diagnostic acceptability (score >= 3), but it was non-inferior when compared to conventional CTU. The mean attenuation value, mean SNR, CNR, and FOM in all pre-defined segments of the urinary tract were significantly higher in the LVLC-CTU group than in the conventional CTU group.Conclusion: The diagnostic acceptability and quantitative image quality of LVLC-CTU with IR are not inferior to those of conventional CTU. Additionally, LVLC-CTU with IR is beneficial because both radiation exposure and total iodine load are reduced.
URI
https://kjronline.org/DOIx.php?id=10.3348/kjr.2018.19.6.1119https://repository.hanyang.ac.kr/handle/20.500.11754/121109
ISSN
1229-6929; 2005-8330
DOI
10.3348/kjr.2018.19.6.1119