Validation of intra- and interobserver reproducibility of shearwave elastography: Phantom study

Title
Validation of intra- and interobserver reproducibility of shearwave elastography: Phantom study
Author
김용수
Keywords
Ultrasound; Elastography; Shearwave elastography; Reproducibility; Phantom; SOLID BREAST MASSES; INITIAL-EXPERIENCE; US ELASTOGRAPHY; LESIONS
Issue Date
2013-07
Publisher
Elsevier Science B.V., Amsterdam.
Citation
Ultrasonics, Vol.53, No.5 [2013], p1039-1043
Abstract
Purpose: To assess intra- and interobserver reproducibility of quantitative ultrasound (US) elastography for breast masses using shearwave elastography (SWE).Materials and methods: The US elastography phantom was used. This training phantom contained 12 masses with dimensions from 6 to 11 mm and varying B-mode (hypo- iso- and hyperechoic) characteristics and viscoelastic properties varying from about 50 to 250 kPa. Four attending radiologists participated as operators and used Supersonic Imaging 15-4 MHz array to acquire images. In each session, B-mode US features were assessed once and US elastography values were assessed twice in a consecutive and distinct manner. A total of three identical sessions of data acquisition were performed at 2-week intervals. For categorical variables such as BI-RADS features of mass on B-mode image and qualitative SWE features, kappa (k) values were calculated by the generalized linear mixed model. For continuous variables, kPa, and ratio, intraclass correlation coefficients (ICCs) were calculated.Results: Interobserver agreement for SWE image similarity was moderate (k = 0.57). With respect to quantitative SWE values, intraobserver ICCs were 0.65, 0.77, 0.92, and 0.91 for maximum elasticity; 0.70, 0.83, 0.94, and 0.94 for mean elasticity; and 0.67, 0.83, 0.92, and 0.92 for elasticity ratio for operators 1, 2, 3, and 4, respectively. Interobserver reproducibility showed good agreement with ICC values of 0.77 for maximum elasticity, 0.82 for mean elasticity, and 0.79 for elasticity ratio.Conclusion: Quantitative US elastography for breast masses was highly reproducible with good agreement across operators. With respect to intraobserver reproducibility, ICC values varied but three out of four operators showed good agreement. (C) 2013 Elsevier B. V. All rights reserved.
URI
http://www.sciencedirect.com/science/article/pii/S0041624X13000310http://hdl.handle.net/20.500.11754/44226
ISSN
0041-624X
DOI
10.1016/j.ultras.2013.01.013
Appears in Collections:
COLLEGE OF MEDICINE[S](의과대학) > MEDICINE(의학과) > Articles
Files in This Item:
There are no files associated with this item.
Export
RIS (EndNote)
XLS (Excel)
XML


qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.

BROWSE