Fixed volume sequential standard addition calibration: Value assignment of impurities in zero gas

Title
Fixed volume sequential standard addition calibration: Value assignment of impurities in zero gas
Author
김기현
Keywords
standard addition; calibration; bias; zero gas; matrix
Issue Date
2017-05
Publisher
ELSEVIER SCIENCE BV
Citation
CHEMOMETRICS AND INTELLIGENT LABORATORY SYSTEMS, v. 164, page. 52-55
Abstract
A fixed volume case of sequential standard addition calibration (S-SAC) for the value assignment of impurities in zero gas is described. A mathematical description of this technique has been derived and has been shown to exhibit a similar systematic bias during the extrapolation process to that seen for other S-SAC cases. The use of S-SAC in the gas phase has demonstrated that variation in the sample volume for S-SAC in the liquid phase is not the generalised cause of the bias experienced during extrapolation. Instead it is the variation in the volume of the original sample with respect to the overall volume of the mixture following the addition of standard. In addition, the requirement to perform a bias correction on the extrapolated values has been discussed and best practice solutions for correction proposed.
URI
https://www.sciencedirect.com/science/article/abs/pii/S0169743917300655?via%3Dihubhttps://repository.hanyang.ac.kr/handle/20.500.11754/114181
ISSN
0169-7439; 1873-3239
DOI
10.1016/j.chemolab.2017.03.012
Appears in Collections:
COLLEGE OF ENGINEERING[S](공과대학) > CIVIL AND ENVIRONMENTAL ENGINEERING(건설환경공학과) > Articles
Files in This Item:
There are no files associated with this item.
Export
RIS (EndNote)
XLS (Excel)
XML


qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.

BROWSE