Analysis of X-ray spectra in 14.5-GHz ECR ion source for optimizing operation conditions

Title
Analysis of X-ray spectra in 14.5-GHz ECR ion source for optimizing operation conditions
Author
이철호
Keywords
X-ray; accelerator; analysis; ECR ion source; X-ray spectra; NaI(Tl) detector; plasma; electron temperature; electron confinement time
Issue Date
2016-06
Publisher
TAYLOR & FRANCIS LTD
Citation
JOURNAL OF NUCLEAR SCIENCE AND TECHNOLOGY, v. 53, NO 12, Page. 2072-2078
Abstract
An electron cyclotron resonance (ECR) ion source operating at 14.5 GHz was developed for the generation of charged ions at the Korea Atomic Energy Research Institute (KAERI). Experiments were carried out to study the plasma inside the ECR ion source by analyzing the X-ray spectra generated by it. The X-ray energy distribution and electron energy inside the plasma chamber are influenced by the status of the heated plasma. That status depends on various operation parameters such as microwave power, injected gas-pressure, and solenoid and trim coil currents. X-ray spectra were recorded to find the correlation between the plasma and the X-rays for variations in the operation parameters. A standard NaI(Tl) detector was used for that purpose. The X-ray energy distribution was studied in the range of 100-500 W for radiofrequency power. The influence of the injected gas pressure and the mirror ratio in the emission of X-rays were analyzed.
URI
https://www.tandfonline.com/doi/full/10.1080/00223131.2016.1195303https://repository.hanyang.ac.kr/handle/20.500.11754/72288
ISSN
0022-3131; 1881-1248
DOI
10.1080/00223131.2016.1195303
Appears in Collections:
COLLEGE OF ENGINEERING[S](공과대학) > NUCLEAR ENGINEERING(원자력공학과) > Articles