리바운드안압계, 비접촉안압계, 골드안 압평안압계의 비교 및 중심각막두께와의 상관성

Title
리바운드안압계, 비접촉안압계, 골드안 압평안압계의 비교 및 중심각막두께와의 상관성
Author
조희윤
Keywords
Central corneal thickness; Goldmann applanation tonometer; Non-contact tonometer; Rebound tonometer
Issue Date
2012-07
Publisher
대한안과학회
Citation
대한안과학회지, 2012, 53(7), P.988-995
Abstract
PurposeTo compare the level of accuracy of intraocular pressure (IOP) measurements of a rebound tonometer (RT) Icare®, and non-contact tonometer (NCT), using Goldmann Applanation tonometer (GAT) as a reference to evaluate the influence of central corneal thickness (CCT) on IOP readings in the Korean population.MethodsIn a prospective study of 273 eyes, IOP was measured with RT, GAT, and NCT and compared to CCT measurements. Patients were assigned to one of 3 groups based on IOP measurements of GAT and 1 of 2 groups based on CCT. The comparison of the IOP values of RT, GAT, and NCT was performed between the IOP and CCT groups, and the differences among tonometers were evaluated.ResultsThe RT showed statistically significant correlation with the GAT compared to the NCT. The CCT was related to RT measurements. The RT values compared to the GAT was underestimated in thin corneas and overestimated in thick corneas.ConclusionsThere was a significant correlation between the RT and the GAT measurements compared with the NCT. However, RT is influenced by CCT and correlates less with GAT in low IOP ranges, suggesting that corneal thickness should be taken into consideration during such measurements.
URI
https://synapse.koreamed.org/DOIx.php?id=10.3341/jkos.2012.53.7.988http://hdl.handle.net/20.500.11754/67260
ISSN
0378-6471
DOI
10.3341/jkos.2012.53.7.988
Appears in Collections:
COLLEGE OF MEDICINE[S](의과대학) > MEDICINE(의학과) > Articles
Files in This Item:
There are no files associated with this item.
Export
RIS (EndNote)
XLS (Excel)
XML


qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.

BROWSE