Comparison of two exhaled nitric oxide analyzers: The NIOX MINO hand-held electrochemical analyzer and the NOA280i stationary chemiluminescence analyzer

Title
Comparison of two exhaled nitric oxide analyzers: The NIOX MINO hand-held electrochemical analyzer and the NOA280i stationary chemiluminescence analyzer
Author
박동원
Keywords
exhaled nitric oxide;; NIOX MINO; nitric oxide analyzer; NOA280i
Issue Date
2012-06
Publisher
BLACKWELL PUBLISHING
Citation
RESPIROLOGY, v. 17, no. 5, Page. 830-834
Abstract
Background and objective: Measurement of the fraction of nitric oxide (FeNO) in exhaled air is useful in the management of asthma. A new hand-held nitric oxide (NO) analyzer, the NIOX MINO, is simple and easy to use in clinical practice. In this study, FeNO values measured using the NIOX MINO were compared with those obtained using a stationary chemiluminescence analyzer, the Sievers NOA280i. Methods: FeNO was measured in 100 adults, using both the NIOX MINO and the NOA280i. Nine (9.0%) of these subjects had asthma. The first acceptable measurement with the NIOX MINO and the mean of two acceptable measurements with the NOA280i were compared. Results: There was a significant correlation between FeNO concentrations measured with the two devices (r = 0.876, P < 0.001). A BlandAltman plot showed a high degree of agreement between the two devices: the mean inter-device difference was 3.3 parts per billion (ppb), and the 95% limits of agreement were -7.0 and 13.6 ppb. In addition, the mean relative difference was 14.5%, with the 95% limits of agreement being -33.7 and 62.7%. The mean value (+/- standard error of the mean) for FeNO as measured with the NIOX MINO (18.8 +/- 0.9 ppb) was significantly lower than that measured with the NOA280i (22.1 +/- 1.2 ppb, P < 0.001). Conclusions: There was a significant correlation, but only moderate agreement, between FeNO values measured with the NIOX MINO and those measured with the NOA280i, with the NIOX MINO values being significantly lower than the NOA280i values. Significant differences in FeNO values obtained with these two NO analyzers should be considered when interpreting the results of FeNO measurements.
URI
http://hdl.handle.net/20.500.11754/57156https://onlinelibrary.wiley.com/doi/full/10.1111/j.1440-1843.2012.02163.x
ISSN
1323-7799; 1440-1843
DOI
10.1111/j.1440-1843.2012.02163.x
Appears in Collections:
COLLEGE OF MEDICINE[S](의과대학) > MEDICINE(의학과) > Articles
Files in This Item:
There are no files associated with this item.
Export
RIS (EndNote)
XLS (Excel)
XML


qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.

BROWSE