216 0

Reliability of a PowerPoint Method for Wear Measurement After Total Hip Arthroplasty A Retrieval Study Using 3-Dimensional Laser Scanning

Title
Reliability of a PowerPoint Method for Wear Measurement After Total Hip Arthroplasty A Retrieval Study Using 3-Dimensional Laser Scanning
Author
양재혁
Keywords
hip; total hip arthroplasty; wear; wear measurement; PowerPoint method
Issue Date
2012-09
Publisher
Churchill Livingstone Inc Medical Publishers
Citation
Journal Of Arthroplasty, 2012, 27(8), P.1530-1537
Abstract
The purpose of this study was to evaluate the reliability of the Power Point (PP) (2007 Version; Microsoft, Redmond, Wash) method for measuring polyethylene liner wear after total hip arthroplasty. Seventeen retrieved polyethylene liners were included in this study. Wear volumes were calculated using the PP, the Dorr and Wan, and 3-dimensional (3D) laser scanning methods. Spearman correlation coefficients for wear volume results indicated strong correlations between the PP and 3D laser scanning methods (range, 0.89-0.93). On the other hand, Spearman correlation analysis revealed only moderate correlations between the Dorr and Wan and 3D laser scanning methods (range, 0.67-0.77). The PP method can be used to monitor linear wear after total hip arthroplasty and could serve as an alternative method when computerized methods are not available.
URI
http://www.arthroplastyjournal.org/article/S0883-5403(11)00705-4/fulltexthttp://hdl.handle.net/20.500.11754/50181
ISSN
0883-5403
DOI
10.1016/j.arth.2011.12.025
Appears in Collections:
COLLEGE OF MEDICINE[S](의과대학) > MEDICINE(의학과) > Articles
Files in This Item:
There are no files associated with this item.
Export
RIS (EndNote)
XLS (Excel)
XML


qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.

BROWSE