277 0

High-fidelity digital recording and playback sphygmomanometry system: device description and proof of concept

Title
High-fidelity digital recording and playback sphygmomanometry system: device description and proof of concept
Author
김인영
Keywords
blood pressure measurement; device validation; digital recording; Korotkoff sounds; sphygmomanometry
Issue Date
2015-10
Publisher
LIPPINCOTT WILLIAMS & WILKINS
Citation
BLOOD PRESSURE MONITORING, v. 20, NO 5, Page. 266-272
Abstract
Objectives This study describes the development of a new digital sphygmocorder (DS-II), which allows the digital recording and playback of the Korotkoff sounds, together with cuff pressure waveform, and its performance in a pilot validation study.Materials and methodsA condenser microphone and stethoscope head detect Korotkoff sounds and an electronic chip, dedicated to audio-signal processing, is used to record high-quality sounds. Systolic and diastolic blood pressure (SBP/DBP) are determined from the recorded signals with an automatic beat detection algorithm that displays the cuff pressure at each beat on the monitor. Recordings of Korotkoff sounds, with the cuff pressure waveforms, and the simultaneous on-site assessments of SBP/DBP were performed during 100 measurements in 10 individuals. The observers reassessed the recorded signals to verify their accuracy and differences were calculated.ResultsThe features of the high-fidelity DS-II, the technical specifications and the assessment procedures utilizing the playback software are described. Interobserver absolute differences (meanSD) in measurements were 0.7 +/- 1.1/1.3 +/- 1.3mmHg (SBP/DBP) with a mercury sphygmomanometer and 0.3 +/- 0.9/0.8 +/- 1.2mmHg with the DS-II. The absolute DS-II mercury sphygmomanometer differences were 1.3 +/- 1.9/1.5 +/- 1.3mmHg (SBP/DBP).ConclusionThe high-fidelity DS-II device presents satisfactory agreement with simultaneous measurements of blood pressure with a mercury sphygmomanometer. The device will be a valuable methodology for validating new blood pressure measurement technologies and devices.
URI
http://pt.wkhealth.com/pt/re/lwwgateway/landingpage.htm;jsessionid=ZJNJz4pGJhPmvpmg2sxplmqvGw8rGVjnphTPYSZgzfFvHgMTpRzb!721960586!181195628!8091!-1?sid=WKPTLP:landingpage&an=00126097-201510000-00006http://hdl.handle.net/20.500.11754/28380
ISSN
1359-5237; 1473-5725
DOI
10.1097/MBP.0000000000000109
Appears in Collections:
COLLEGE OF MEDICINE[S](의과대학) > MEDICINE(의학과) > Articles
Files in This Item:
There are no files associated with this item.
Export
RIS (EndNote)
XLS (Excel)
XML


qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.

BROWSE