Development of Multichannel Ellipsometry with Synchronously Rotating Polarizer and Analyzer

Title
Development of Multichannel Ellipsometry with Synchronously Rotating Polarizer and Analyzer
Author
안일신
Keywords
multichannel; spectroscopic; ellipsometry; calibration; rotating
Issue Date
2002-06
Publisher
IOP Publishing Ltd
Citation
Japanese Journal of Applied Physics, v. 41, no. 6R, page. 3978-3980
Abstract
Rotating polarizer and analyzer ellipsometry was developed with a multichannel detector. In this system, the transmission axes of polarizer and analyzer rotated in phase using flexible shaft for mechanical connection. With this configuration, auto-calibration for the azimuths of polarizer and analyzer was achieved. Thus, experimental error related to the calibration could be avoided and spectroscopic ellipsometry in the simplest operation principle was realized. Data acquisition time for 512 points ellipsometry spectra was 3 sec over the photon energy range of 1.5 to 4.5 eV.
URI
https://iopscience.iop.org/article/10.1143/JJAP.41.3978/metahttps://repository.hanyang.ac.kr/handle/20.500.11754/157016
ISSN
0021-4922; 1347-4065
DOI
10.1143/jjap.41.3978