Comparison of the Bone Harvesting Capacity of an Intraoral Bone Harvesting Device and Three Different Implant Drills
- Title
- Comparison of the Bone Harvesting Capacity of an Intraoral Bone Harvesting Device and Three Different Implant Drills
- Author
- 한지영
- Keywords
- MAXILLARY SINUS AUGMENTATION; PARTICLE-SIZE; AUTOGENOUS BONE; RIDGE AUGMENTATION; GRAFT; OSTEOTOMY; COLLECTORS; MARROW; VOLUME
- Issue Date
- 2017-12
- Publisher
- HINDAWI LTD
- Citation
- BIOMED RESEARCH INTERNATIONAL, Article no. 7819080
- Abstract
- The aim of the present study was to compare bone-collecting capacity of bone harvesting device and minimally irrigated low-speed drilling using three implant systems. One bone harvesting device and three commercially available drill systems were compared using the osteotomies on bovine rib bones. The amount of the collected bone particle and particle size (<500 mu m: small, 500-1000 mu m: medium, and >1000 mu m: large) were measured. Total wet (1.535 +/- 0.232 mL) and dry volume (1.147 +/- 0.425 mL) of the bone particles from bone harvesting device were significantly greater than three drill systems (wet volume: 1.225 +/- 0.187-1.27 +/- 0.29 mL and dry volume: 0.688 +/- 0.163-0.74 +/- 0.311 mL) (p <0.05). In all groups, the amount of large sized particles in wet and dry state was the greatest compared to that of medium and small particles. The dry weight of the bone particles showed the same tendency to volumetric measurement. In conclusion, total bone particles and large sized particles (>1000 mu m) were harvested significantly greater by bone harvesting device than minimally irrigated low-speed drilling. The composition of particle size in all harvesting methods was similar to each other.
- URI
- https://www.hindawi.com/journals/bmri/2017/7819080/https://repository.hanyang.ac.kr/handle/20.500.11754/116681
- ISSN
- 2314-6133; 2314-6141
- DOI
- 10.1155/2017/7819080