Measurements and prediction of the ambient dose rate from patient receiving radioiodine administration after thyroid ablation

Title
Measurements and prediction of the ambient dose rate from patient receiving radioiodine administration after thyroid ablation
Author
이재기
Issue Date
2012-08
Publisher
OXFORD UNIV PRESS, GREAT CLARENDON ST, OXFORD OX2 6DP, ENGLAND
Citation
RADIATION PROTECTION DOSIMETRY, 권: 151, 호: 1, 페이지: 158-161
Abstract
Empirical equations were proposed for the prediction of the ambient dose equivalent rates from patients administered with radioiodine for the treatment of thyroid cancers. Ambient dose equivalent rates were measured for 59 patients who received high-dose I-131 treatment after thyroid ablation at different times after the administration. An ion chamber was used to measure the dose equivalent rates at 1 m away from the body and on contact to the thyroid of the patients. The resulting equations for estimating dose equivalent rates at 1 m from the body and on contact to the thyroid are, respectively, dot H(body)(1)0.236e(0.0501t) (mSv h(1) GBq(1)) and dot H(thy)(c)2.676e(0.0443t) (mSv h(1) GBq(1)). The effective half-times in total body appeared to be 13.86 h.
URI
http://academic.oup.com/rpd/article/151/1/158/1598993http://hdl.handle.net/20.500.11754/41008
ISSN
0144-8420
DOI
10.1093/rpd/ncr456