26개 수련병원 응급의료센터 내 비치된 직접후두경 밝기의 적절성

Title
26개 수련병원 응급의료센터 내 비치된 직접후두경 밝기의 적절성
Other Titles
Original Articles : Adequacy of Direct Laryngoscope Brightness in 26 Emergency Rooms
Author
임태호
Keywords
응급의료센터; 직접후두경; Laryngoscopes; Emergency medicine; Light
Issue Date
2011-10
Publisher
대한응급의학회
Citation
대한응급의학회지, Vol. 22 No. 5, 466-470
Abstract
Purpose: To determine adequacy of brightness and proper brightness of the direct laryngoscope in emergency departments. Methods: We investigated blade and handle in 26 hospitals in Seoul and made 110 combination. We excluded malfunction and non-fixed combination, and measured the lux of each combination. We divided results into two groups (bulb type and fiber type) in two ways (adult type and pediatric type). Results: In the 110 combinations, the median lux value was 836.25. The number of laryngoscopes in which brightness exceeded International Organization for Standardization (ISO) recommended value of the 700 was 67. There was no statistically significant difference in median lux between fiber type and bulb type (847.5 vs 802.0; p=0.870), while the brightness of the pediatric type was brighter than the adult type. A questionnaire of laryngoscope management revealed no regular basis of use, and complete absence of awareness of brightness criteria. Conclusion: An improved laryngoscope managing system is essential including regular criteria for laryngoscope management.
URI
http://kiss.kstudy.com/thesis/thesis-view.asp?key=2948010http://hdl.handle.net/20.500.11754/36869
ISSN
1226-4334
Appears in Collections:
COLLEGE OF MEDICINE[S](의과대학) > MEDICINE(의학과) > Articles
Files in This Item:
There are no files associated with this item.
Export
RIS (EndNote)
XLS (Excel)
XML


qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.

BROWSE