113 0

A single-walled carbon nanotube thin film-based pH-sensing microfluidic chip

Title
A single-walled carbon nanotube thin film-based pH-sensing microfluidic chip
Author
성기훈
Issue Date
2014-03
Publisher
ROYAL SOC CHEMISTRY
Citation
ANALYST, v. 139, NO. 8, Page. 2011-2015
Abstract
A novel microfluidic pH-sensing chip was developed based on pH-sensitive single-walled carbon nanotubes (SWCNTs). In this study, the SWCNT thin film acted both as an electrode and a pH-sensitive membrane. The potentiometric pH response was observed by electronic structure changes in the semiconducting SWCNTs in response to the pH level. In a microfluidic chip consisting of a SWCNT pH-sensing working electrode and an Ag/AgCl reference electrode, the calibration plot exhibited promising pH-sensing performance with an ideal Nernstian response of 59.71 mV pH(-1) between pH 3 and 11 (standard deviation of the sensitivity is 1.5 mV pH(-1), R-2 = 0.985). Moreover, the SWCNT electrode in the microfluidic device showed no significant variation at any pH value in the range of the flow rate between 0.1 and 15 mu l min(-1). The selectivity coefficients of the SWCNT electrode revealed good selectivity against common interfering ions.
URI
https://pubs.rsc.org/en/content/articlelanding/2014/AN/c3an02195ehttps://repository.hanyang.ac.kr/handle/20.500.11754/183546
ISSN
0003-2654;1364-5528
DOI
10.1039/c3an02195e
Appears in Collections:
COLLEGE OF ENGINEERING SCIENCES[E](공학대학) > BIONANO ENGINEERING(생명나노공학과) > Articles
Files in This Item:
There are no files associated with this item.
Export
RIS (EndNote)
XLS (Excel)
XML


qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.

BROWSE