Comparison of EMG signals recorded by surface electrodes on endotracheal tube and thyroid cartilage during monitored thyroidectomy

Title
Comparison of EMG signals recorded by surface electrodes on endotracheal tube and thyroid cartilage during monitored thyroidectomy
Author
지용배
Keywords
Electromyography recording; Intraoperative neural monitoring; Recurrent laryngeal nerve; Vocal cord paralysis; Thyroid surgery
Issue Date
2017-10
Publisher
ELSEVIER TAIWAN
Citation
KAOHSIUNG JOURNAL OF MEDICAL SCIENCES, v. 33, no. 10, page. 503-509
Abstract
A variety of electromyography (EMG) recording methods were reported during intraoperative neural monitoring (IONM) of recurrent laryngeal nerve (RLN) in thyroid surgery. This study compared two surface recording methods that were obtained by electrodes on endotracheal tube (ET) and thyroid cartilage (TC). This study analyzed 205 RLNs at risk in 110 patients undergoing monitored thyroidectomy. Each patient was intubated with an EMG ET during general anesthesia. A pair of single needle electrode was inserted obliquely into the TC lamina on each side. Standard IONM procedure was routinely followed, and EMG signals recorded by the ET and TC electrodes at each step were compared. In all nerves, evoked laryngeal EMG signals were reliably recorded by the ET and TC electrodes, and showed the same typical waveform and latency. The EMG signals recorded by the TC electrodes showed significantly higher amplitudes and stability compared to those by the ET electrodes. Both recording methods accurately detected 7 partial loss of signal (LOS) and 2 complete LOS events caused by traction stress, but only the ET electrodes falsely detected 3 LOS events caused by ET displacement during surgical manipulation. Two patients with true complete LOS experienced temporary RLN palsy postoperatively. Neither permanent RLN palsy, nor complications from ET or TC electrodes were encountered in this study. Both electrodes are effective and reliable for recording laryngeal EMG signals during monitored thyroidectomy. Compared to ET electrodes, TC electrodes obtain higher and more stable EMG signals as well as fewer false EMG results during IONM. Copyright (C) 2017, Kaohsiung Medical University.
URI
https://onlinelibrary.wiley.com/doi/full/10.1016/j.kjms.2017.06.014https://repository.hanyang.ac.kr/handle/20.500.11754/115814
ISSN
1607-551X
DOI
10.1016/j.kjms.2017.06.014
Appears in Collections:
COLLEGE OF MEDICINE[S](의과대학) > MEDICINE(의학과) > Articles
Files in This Item:
Biological function integrated prediction of severe radiographic progression in rheumatoid arthritis a nested case control study.pdfDownload
Export
RIS (EndNote)
XLS (Excel)
XML


qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.

BROWSE