2-dimensional shear wave elastography: Interobserver agreement and factors related to interobserver discrepancy

Title
2-dimensional shear wave elastography: Interobserver agreement and factors related to interobserver discrepancy
Author
김용수
Keywords
LIVER FIBROSIS; TRANSIENT ELASTOGRAPHY; RHEUMATOID-ARTHRITIS; HEPATITIS-C; STIFFNESS; RELIABILITY; REPRODUCIBILITY; ELASTICITY; ACCURACY; ARFI
Issue Date
2017-04
Publisher
PUBLIC LIBRARY SCIENCE
Citation
PLOS ONE, v. 12, no. 4, Article no. e0175747
Abstract
Purpose To evaluate the interobserver reproducibility of two-dimensional shear wave elastography (2D-SWE) in measuring liver stiffness (LS) and to investigate factors related to liver 2D-SWE. Materials and methods A prospective study was conducted between August 2011 and August 2012 in rheumatoid arthritis patients who had been treated with methotrexate. Interobserver reproducibility of 2D-SWE was evaluated, and the relationship between interobserver difference in LS and related factors was analyzed using linear regression analyses. We considered age, sex, alanine transaminase, cholesterol, body mass index (BMI), and waist circumference as clinical factors, and the mean value of standard deviation (SDM), its difference between two examiners, mean diameter of the regions of interest (ROIM), and its difference in the elasticity map as investigation factors. The cut-off value for significant factors to predict interobserver discrepancies in LS-based fibrosis stage was also inspected. Results In total, 176 patients were enrolled. The intraclass correlation coefficient between the two examiners was 0.784. In the univariate analysis, SDM and ROIM were independently associated with interobserver differences in LS as well as BMI, waist circumference, and the difference of ROI, but SDM and ROIM were the only ones significantly related in multivariate analysis (p<0.001 and p = 0.021, respectively). The best cut-off value for SDM in predicting interobserver discrepancy in LS-based fibrosis stage was 1.4. Conclusions Interobserver reproducibility of 2D-SWE for measuring LS was good and SDM was the most significantly associated factor with interobserver differences in LS and interobserver discordance in LS-based fibrosis stage.
URI
https://journals.plos.org/plosone/article?id=10.1371/journal.pone.0175747https://repository.hanyang.ac.kr/handle/20.500.11754/113662
ISSN
1932-6203
DOI
10.1371/journal.pone.0175747
Appears in Collections:
COLLEGE OF MEDICINE[S](의과대학) > MEDICINE(의학과) > Articles
Files in This Item:
2-dimensional shear wave elastography Interobserver agreement and factors related to interobserver discrepancy.pdfDownload
Export
RIS (EndNote)
XLS (Excel)
XML


qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.

BROWSE